Repositorio Institucional
Please use this identifier to cite or link to this item:
https://saber.ucv.ve/handle/10872/14484
|
| Title: | Optimization of experimental parameters for the determination of mercury by MIP:AES |
| Authors: | Murillo, Miguel Carrión, Nereida Chirinos, José Gammeiro, Alexis Fassano, Edmondo |
| Keywords: | Microwave Plasma MIP Surfatron Mercury determination |
| Issue Date: | 2001 |
| Publisher: | Atlanta |
| Citation: | Talanta 54 (2001) 389–395 |
| Abstract: | A system using a microwave induced plasma, as an emission source was optimized for total mercury determinations. The system contains a flow injection section in which mercury is reduced and deposited on a gold:platinum collector, and carried to the emission source. The microwave was generated in a Surfatron cavity. An experimental design was applied to optimize parameters concerning plasma conditions, amalgamation, desorption and mercury vapor generation. The limit of detection of the method was 14 pg ml 1 of mercury, using peak area mode while LDs was 1.3 pg ml 1 using peak height mode, with a 2% RSD and a linear dynamic range of almost three orders of magnitude. The accuracy was assessed using a certified reference material of atmospheric particulate National Institute of Standards and Technology (NIST) 2704. No significant difference at 95% confidence level was observed between the certified value and our results. |
| URI: | http://hdl.handle.net/10872/14484 |
| ISSN: | 0039- 9140 |
| Appears in Collections: | Artículos Publicados |
Files in This Item:
|
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.