Título : | Investigation of the direct injection high efficiency nebulizer for axially and radially viewed inductively coupled plasma atomic emission spectrometry |
Autor : | Su-Ann E., O’Brien, Chirinos, José Kaveh, Jorabchi Kaveh, Kahen, Michelle, E. Cree Montaser, Akbar |
Fecha de publicación : | 2003 |
Resumen : | A direct injection high efficiency nebulizer (DIHEN) is explored for introduction of the sample aerosol into the
central channel of the axially and radially viewed inductively coupled plasma (ICP) of a commercial ICP
atomic emission spectrometer (ICPAES). The DIHEN is a micro-nebulizer that requires very low solution
uptake rates (1–100 mL min21) and nebulizer gas flow rates (v0.2 L min21) compared to conventional
nebulizer-spray chamber arrangements (y1.0 mL min21 and y1.0 L min21, respectively). Signal-tobackground
ratios (SBRs), detection limits, and precision of the DIHEN are comparable or superior to those
of the conventional sample introduction system, but the Mg II 280.270/Mg I 285.213 nm ratios are lower with
the DIHEN, indicating that the DIHEN is more susceptible to matrix effects than the conventional
nebulization system, for both the axial and radial ICPAES systems. Matrix effects are further investigated by
comparing intensity ratios with and without 0.1% and 0.5% Na for several spectral lines having energy sum
ranging from 7.93 to 14.79 eV. Replacement of Ar with Ar–O2 and Ar–N2 mixtures in the outer gas flow of the
plasma improves SBRs and Mg II 280.270/Mg I 285.213 nm ratios of the DIHEN, and reduces matrix effects.
By reducing solution uptake rate from 60 to 30 mL min21, matrix effect is also reduced. Operation of the radial
instrument at 1700 W reduces matrix effect compared to the effect noted for the axial instrument at 1500 W.
Finally, the utility of the technique in practical ICPAES studies is demonstrated using a custom made organometallic
standard for As, Hg, and Pb in xylene. |